6.4.4.4 Managing Reference Weights
For best results begin with the highest nominal value when entering data.
Sufficient reference weights must be available to ensure that the weight differences on substitution do not amount to more than 100 g.
For reasons of traceability, only certified, calibrated reference weights should be entered.
6.4.4.5 Managing Operators
Enter the details of the operators who have been trained to work with the Pycnometer.
A repeatability add-on may be entered if different standard deviations typically occur for different operators when handling the
measurement sequences.
6.4.4.6 Managing Pressure, Temperature and Relative Humidity Sensors
Note: Ensure that the algebraic sign is correct when entering the correction value. The correction value is added (not subtracted) when
analyzing the measurements entered as the "readout" values.
For reasons of traceability, only sensors supported by a calibration certificate should be entered.