Measuring range °F
-58 to +68 °F
+68 to +932 °F
+932 to +1832 °F
+1832 to +2192 °F
Measuring area °C
-50 to 0 °C
0 to +1370 °C
Measuring range °F
-58 to +32 °F
+32 to +2498 °F

b) Degree of Emission of Different Surfaces

The emission degrees listed in the table are approximates. Various parameters, such as geometry and surface qual-
ity, may influence the emission degree of an object.
The meter is delivered with a pre-set emission degree of 0.95. It matches for most non-metal materials. The IR
measuring method is only conditionally suitable for blank metals and requires
special surface treatment (e.g. matte insulating tape, etc.) or pre-setting of the corresponding emission degree.
Surface
Aluminium, blank
Aluminium, oxidised
Asphalt
Concrete
Ice
Iron oxide
Colour
Plaster
Glass/porcelain
Rubber, black
Wood
50
Accuracy
±9 °F
±1.5% or
±3.6 °F
±2,5 %
±3,0 %
Contact temperature measurement type -K
Accuracy
±3 °C
±(1% + 2 °C)
Accuracy
±3.6 °F
±(1% + 3 °F)
Emission Degree
0,04
0,30
0,90 - 0,98
0,94
0,96 - 0,98
0,78 - 0,82
0,93
0,80 - 0,90
0,92 - 0,94
0,94
0,94
Reproducibility
±2.7 °F
±0.5% or
±0.9 °F
±0.5% or
±0.9 °F
±1,0 %
Surface
Copper, oxidised
Brass, oxidised
Paints, matte
Foods
Human skin
Plastic
Paper
Sand
Textiles
Water
Brick, plaster
Emission Degree
0,60
0,50
0,93 - 0,97
0,93 - 0,98
0,98
0,94
0,97
0,90
0,90
0,92 - 0,96
0,93 - 0,96