15. Remote commands

The data request is done by the sending of the character ASCII <?>, followed by <CR> (carriage return). The instru-
ment also answers in the same way during the detection of an impulse pedals/automat.
The other commands are shown in the following table:
AOFF 0 / AOFF 1
AOFF?
BAT?
ID ?
IN
KEY?
KEY 0 / KEY 1
LCAL?
LCAL DD.MM.[YY]YY
MM
NCAL?
NCAL DD.MM.[YY]YY
NUM ?
Deactivates/activates the automatic switch-off
Forwards the state of the automatic switch-off
Sends the status of the battery (BAT1=ok or BAT0=low)
Sends the instrument identification (SY276)
Activates the Inch unit
Sends keyboard status (KEY0 or KEY1)
Locks/unlocks the keyboard (except data output)
Sends the date of the last calibration (DD.MM.YYYY)
Sets the date of the last calibration
Activates the millimeter unit
Sends the date of the next calibration (DD.MM.YYYY)
Sets the date of the next calibration
Sends the instrument number
18