H82.0.01.6C-03

15 Specifications

Display range
Number of channels
Suitable probes
Available ranges/resolutions
Connections
sensor
output /
ext. supply
Display units
User unit
Measuring frequency
Accuracy
Working conditions
Storage conditions
Display
Correction function
Additional functions
Averaging
Housing
protection class IP65 and IP67
dimensions
L*W*H [mm]
Power supply
current
consumption
Battery-change indicator
Auto-off-function
EMC
Operating Manual GMH 5130
_____________________________________________________
maximal -19999...19999 digit, depends on used sensor
1
GMSD ... - K51 and MSD ... sensors with connection cable MSD-K51
ranging from -1.999 ... 2.500 mbar / 0.001 mbar to 0.0 ... 1000 bar / 1 bar
7-pole bayonet connector
4-pole connector for serial interface and supply (USB Adapter USB 5100)
depends on meas. range selection (depends on connected sensor):
mbar, bar, Pa, kPa, MPa, mmHg, inHg, PSI, mH
freely adjustable
4 / s or 1000 / s
± 0.1 % FS ± 1 digit
-25 ... 50 °C; 0 ... 95 %RH (non condensing)
-25 ... 70 °C
4 ½ digit 7-segment display, illuminated (white)
offset / slope (via menu),
if offset/slope value not equal zero: "corr" is displayed
min- / max- / hold- function
adjustable, 1 ... 120 seconds
160 * 86 * 37 incl. protective covering,
approx. 250 g incl. batteries and protective covering
2*AAA-batteries (included in delivery)
2.0 mA (at Out = Off, equivalent to 500 h),
illumination ~10mA (turns off automatically)
automatically if battery exhausted "bAt", warning "bAt" flashing
Device will be automatically switched off if no key is pressed/no interface
communication takes place for the time of the power-off delay.
The power-off delay can be set to values between 1 and 120 min.; it can be
completely deactivated.
The device corresponds to the essential protection ratings established in the
Regulations of the Council for the Approximation of Legislation for the member
countries regarding electromagnetic compatibility (2004/108/EG)
Additional fault: <1%
_____________________________________________________________________________
O
2
page 13 of 14