13 Error and System Messages

Display
Description
No sensor connected
Connected sensor or device defective
Err.9
or
Value extremely out of measuring range
Battery empty
No display or
confused
Mains operation: wrong voltage or polarity
characters,
System error
device does not
react on keypress
Device defective
Err.1
Measured value above allowable range
Sensor defective
Err.2
Measured value below allowable range
Sensor defective
Display range overflow
Err.3
Display range underflow
Err.4
Value could not be calculated
Er.11
Calculation overflow happened
Err.7
System error
Sensor not present / recognised
– – – –
could not calculate value
If "bAt" is flashing, the battery will be exhausted soon. Further measurements are possible for short time.
If "bAt" is displayed continuously the battery is ultimately exhausted and has to be replaced. Further
measurements aren‟t possible any more.

14 Reshipment und Disposal

14.1 Reshipment

All devices returned to the manufacturer have to be free of any residual of measuring media
and/or other hazardous substances. Measuring residuals at housing or sensor may be a risk for
persons or environment
_DANGER
Use a adequate transport package for reshipment, especially for fully functional devices. Please
make sure that the device is protected in the package by enough packing materials.

14.2 Disposal instructions

Batteries must not be disposed in the regular domestic waste but at the designated collecting points.
The device must not be disposed in the unsorted municipal waste! Send the device directly to us
(sufficiently stamped), if it should be disposed. We will dispose the device appropriate and environmentally
sound.
Operating Manual GMH 5130
_____________________________________________________
What to do?
Switch off device and connect sensor
If 2nd sensor available, check if device is ok. Return
defective device/sensor to manufacturer for repair
Check: pressure not within sensor range?
Replace battery
Check power supply, replace it when necessary
Disconnect battery and power supplies, wait shortly,
then reconnect
Return to manufacturer for repair
Check: pressure not within sensor range?
-> measuring value to high!
Return to manufacturer for repair
Check: pressure not within sensor range?
-> measuring value to low!
Return to manufacturer for repair
Check: value above 19999 -> to high to be displayed
Check: value below -19999 (Tara?) -> to low
Choose different unit
Choose different unit
Return to manufacturer for repair
reconnect sensor,
during logging: stop the logger and restart it
suitable sensor / unit combination necessary
_____________________________________________________________________________