Deactivating the automatic shut off
In order to save the batteries the device is equipped with an auto-
matic power off function (automatic shut off). If the device is not
operated for more than 30 minutes it will automatically shut off. The
automatic shut off is activated when the hand thermometer is
switched on. When activated, the automatic shut off is signalled in
the LCD display with the automatic power off symbol "
".
To deactivate this function switch the hand thermometer off. Press
and hold the Hold button (6) while switching the hand thermometer
back on. The automatic shut off is now deactivated and the auto-
matic power off symbol "
" will be extinguished.
Calibration
The hand thermometer is calibrated in the factory. For precise mea-
surements the hand thermometer can be set to the respective tem-
perature sensor. Hold the temperature sensor in a container of icy
water (0°C). Check the hand thermometer's display after approxi-
mately one minute. If it does not read 0°C it can be correctly set by
turning the offset calibration screw (9). For more precise measure-
ments the hand thermometer can be set with an optional thermocal-
ibrator.

Maintenance and cleaning

Clean the casing occasionally with a dry cloth. Never use liquids nor
cleaning agents. Maintenance or repairs must be carried out by
qualified personnel.
Remove batteries from the device when it is not to be used for an
extended period.