Recommended separation distances
Recommended separation distances between portable and mobile RF communications equipment and the
EQUIPMENT or SYSTEM – for EQUIPMENT or SYSTEM that are not LIFESUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the Blood pressure
monitor BM 27.
The Blood pressure monitor BM 27 is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Blood pressure monitor BM 27 can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
Blood pressure monitor BM 27 as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter
(W)
0,01
0,1
1
10
100
Separation distance according to frequency of transmitter (m)
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.1167
0.2334
0.3689
0.7378
1.1667
2.3334
3.6893
7.3786
11.6667
23.3334