10. Device error and troubleshooting
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «Err 3», is displayed.
Error
Description Potential cause and remedy
«Err 1»
Signal too
The pulse signals on the cuff are too
weak
weak. Re-position the cuff and repeat
AQ
the measurement.*
«Err 2»
Error signal During the measurement, error signals
were detected by the cuff, caused for
AP-B
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«Err 3»
Abnormal
An adequate pressure cannot be gener-
cuff pres-
ated in the cuff. A leak may have
AP
sure
occurred. Check that the cuff is
correctly connected and is not too
loose. Replace the batteries if neces-
sary. Repeat the measurement.
Make sure the cuff connector is
securely inserted into the cuff socket of
your blood pressure monitor. A distinct
«CLICK» must be heard when fully
inserted.
«Err 5» Abnormal
The measuring signals are inaccurate
result
and no result can therefore be
displayed. Read through the checklist
for taking a reliable measurement and
then repeat the measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high
pressure too
(over 299 mmHg) OR the pulse is too
high
high (over 200 beats per minute). Relax
for 5 minutes and repeat the measure-
ment.*
«LO»
Pulse too
The pulse is too low (less than 40 beats
low
per minute). Repeat the measurement.*
Microlife BP B1 Smart
Error
Description Potential cause and remedy
®
Bluetooth
Bluetooth
symbol BN
the device Bluetooth
blinks
app on the smartphone. Wait for 1
rapidly
minute, open the app on the smart-
phone and manually activate Blue-
tooth
tooth
«Err bt» Bluetooth
®
Bluetooth
self check
Contact your local Microlife distributor.
error
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
11.Device maintenance and disposal
Cleaning the device
The device can be cleaned when necessary (e.g., between uses
by different patients).
Use a soft cloth, dry or wet with detergent, to gently wipe the
exterior of the device remove dust or stains.
Cleaning the cuff
Use a soft cloth, dry or wet with mild detergent, to carefully wipe
the cuff to remove dust or stains.
Caution: Do not wash the cuff in a washing machine or
dishwasher!
Cleaning the AC adapter
Clean the AC adapter with a dry cloth.
Storage
When not in use:
 Disconnect the cuff and parts from the device.
 Keep the device and accessories in a dry, cool place away from
sunlight, with ambient conditions within the temperature and
humidity ranges described in the «Specifications and compli-
ance» section.
 Remove the batteries from the device if the device will not be
used for an extended period.
®
connection error. Turn off
®
and close the
®
on the device, to re-try Blue-
®
connection and data transfer.
®
is malfunctioning.
9
EN
Inhaltsverzeichnis
loading

Inhaltsverzeichnis