9. Specifications

Type:
BC 16
Measuring method:
non-invasive, oscillometric blood
pressure measurement at the wrist, for
wrist circumferences
of 13.5–22 cm
Display range:
0–299 mmHg
Measuring range:
systolic: 40–250 mmHg / diastolic:
40–250 mmHg /
pulse: 40–199/min
Accuracy of the
systolic ±3 mmHg /
pressure reading:
diastolic: ±3 mmHg /
pulse: ±5% of the displayed value
Measuring uncertainty:
(max. permissible standard deviation
according to clinical testing: systolic: 8
mmHg / diastolic:
8 mmHg
Power supply:
2 x 1.5 V Micro
(Alkaline Type LR 03)
Classifi cation:
Application part Type BF
Ambient conditions:
+10 °C – +40 °C max. 40-85%
atmospheric humidity
Storage/transport:
-10 °C – +60 °C max. 10-90%
atmospheric humidity
Key to symbols:
Application part Type BF
Important! Read the operating
instructions
These specification are subject to change without notice for
purpose of improvement.
• This unit is in line with European Standard EN60601-1-2
and is subject to particular precautions with regard to
electromagnetic compatibility (EMC). Please note that
portable and mobile HF communication systems may
interfere with this instrument. For more details, please
contact customer service at the address indicated.
• The instrument is in line with the EU Medical Products
Directive 93/42/EC, the German medical products act
and European standards EN1060-1 (non-invasive blood
pressure monitors, Part 1: general requirements) and
EN1060-3 (non-invasive blood pressure monitors, Part
3: Supplementary Requirements for Electromechanical
Blood Pressure Measurement Systems).
• If you are using the instrument for commercial or econo-
mic purposes, you must check measurement precision
regularly in accordance with the "Operators' Ordinance
on Medical Products". Even in the case of private use, we
recommend checking measurement precision every two
years at the manufacturers.