7.2. GENERAL CHARACTERISTICS

Reference guidelines
Safety:
EMC:
Insulation:
Pollution degree:
Measurement category:
Display
Characteristics:
Over range indication:
Display rating:
Conversion mode:
Power supply
Battery type:
Low battery indication:
Battery life:
Protection fuses:
Mechanical characteristics
Dimensions (L x W x H):
Weight (including battery):
Mechanical protection:
Environmental conditions
Reference temperature:
Operating temperature:
Operating humidity:
Storage temperature:
Storage humidity:
Max height of use:
This instrument satisfies the requirements of Low Voltage Directive
2014/35/EU (LVD) and of EMC Directive 2014/30/EU
This instrument satisfies the requirements of European Directive
2011/65/EU (RoHS) and 2012/19/EU (WEEE)
7.3. ACCESSORIES

7.3.1. Standard accessories

Couple of test leads
Batteries (not fitted)
Quick reference guide
30
IEC/EN61010-1
IEC/EN61326-1
double insulation
2
CAT III 600V to ground
3½ LCD (2000counts), decimal point
unit symbol, backlight
symbol "OL" on the display
2 times/s
Mean value
2x1.5V batteries type AAA LR03
symbol "
" on the display
ca 30h (backlight ON)
ca 130h (backlight OFF)
F10A/600V, 5 x 20mm (10A input)
F200mA/600V, 5 x 20mm (mAA input)
125 x 70 x 40mm (5 x 3 x 2in)
150g (5 ounces)
IP40
18°C  28°C (64°F ÷ 82°F)
0°C ÷ 50°C (32°F ÷ 122°F)
<70%RH
-20°C ÷ 60 °C (-4°F ÷ 140°F)
<80%RH
2000m (6.562ft)
HT15D