RID14
4.3.1
Turning the display
2.
1.
1
3
Field indicator, 4 display positions, can be fitted in 90° stages
The display unit can be turned in 90 ° stages.
1.
First remove the cover clamp (1) and the housing cover (2).
2.
Then disconnect the display (3) from the electronics unit (4).
3.
Turn the display to the desired position and then fit it onto the electronics unit.
4.3.2
Mounting directly on the wall
Proceed as follows to mount the device directly on the wall:
1.
Drill 2 holes
2.
Fit the device on the wall with 2 screws (⌀5 mm (0.2 in)).
4.3.3
Pipe mounting
The mounting bracket is suitable for pipes with a diameter between 1.5" to 3.3".
The additional mounting plate must be used for pipes with a diameter between 1.5" to 2.2".
The mounting plate is not needed for pipes with a diameter between 2.2" to 3.3".
Proceed as follows to mount the device on a pipe:
Endress+Hauser
3
2
Pos. 1
3.
Pos. 2
90°
4
Pos. 3
90°
Installation
Pos. 4
90°
A0023724
31