medicap Precise 6000 MI Gebrauchsanweisung Seite 71

Sauerstoffkonzentrator
Vorschau ausblenden Andere Handbücher für Precise 6000 MI:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Précise 6000 MS / Précise 6000 MSI
Device has overheated
1) Device was positioned in
the wrong place,
2) Coarse dust filter is
clogged
Device switches
automatically to error mode
1) Coarse dust filter is
clogged
2) Suction filter is clogged
3) Device defective
Device automatically
switches to error mode
Device defective
(column pack too damp)
Device automatically
switches to error mode
Tube connection valve -
column leaking
Device automatically
switches to error mode
Ionising head not screwed
on correctly
Rev.D 2024-10
Allow device
to
cool down for
30 min, set up correctly and
switch on again, clean the
filter fleece and switch on the
device again
Call Service if the message is
repeated
Clean the filter fleece and
switch it on again
Call Service if the message is
repeated
Call Service
Call Service
Check ionisiation head,
screw on correctly and hand-
tight
71