Generally, the use of disposable bubble humidifiers with sterile water is recommended for molecular oxygen
inhalation. For reusable bubble humidifiers, the humidifier bottles must be cleaned in accordance with the
manufacturer's instructions.
The desired molecular mode Standard or ionisation mode Standard can only be preset by a service technician:
Note
The desired molecular mode Standard or ionisation mode Standard can only be preset by a
service technician:
•
Standard: Oxygen supply without timer function. An application duration cannot be set.
Note
Molekular mode therapy or ionisation mode therapy:
Application duration can be set with an acoustic signal after the set time has elapsed.
3.4.1 Overview of molecular mode therapy
The application duration can only be set before starting the application; once the application has been
successfully started, the application duration can no longer be set.
•
Maximum 2 hours
•
Minimum 5 minutes
•
Adjustments possible at 5-minute intervals
•
The user can pause and stop the application using the symbols
Précise 6000 MS / Précise 6000 MSI
and
Rev.D 2024-10
.