3.4.5 Night mode
Night mode can be set on the device. The time after which the device switches to night mode can be set between
1 and 10 minutes. The time setting must be confirmed with
.
If night mode is active, the display no longer lights up. Night mode can be deactivated by simply touching the
display.
3.5 Switching off
Press ON/OFF button for approx. 3 seconds. The display switches off and an acoustic signal sounds. If there is
no more operation on the same day, the humidifier bottle used must be cleaned. Do not store the water (germ
formation). If the device is not to be used for a longer period of time, the plug must be removed from the socket.
3.6 Oxygen Concentration Supervision
The oxygen concentration in the product gas is permanently monitored using a wear-free, ultrasound-based
oxygen sensor. If limit values are not reached, these are shown on the display. The sensor does not need to be
recalibrated.
3.7 Service & Diagnosis
The Précise 6000 MI / MSI oxygen concentrator features comprehensive self-diagnostics.
To access the "System settings", tap on the area marked in red. After exiting the system settings automatically or
manually, the diagnostic mode is executed.
After approx. 30 seconds of inactivity, the software automatically exits the system settings.
Précise 6000 MS / Précise 6000 MSI
Rev.D 2024-10
66