4 Malfunctions and their elimination
Failure symptom
Précise 6000 MS / Précise 6000 MSI
Cause of error
No power or device fuse
defective
Device switches off
automatically
Oxygen concentration
< 82%
Oxygen concentration
< 72%
Device automatically
switches to error mode
Oxygen tube to the mask
kinked
Rev.D 2024-10
Remedy
Check whether the power
cable is connected correctly,
check the device fuse or main
fuse, see picture „Detail view
mains socket"
Call Service
Reduce flow
Call Service if the message is
repeated
Switch device off and on
again
Call Service if the message is
repeated
Check the tube or mask