Maintenance and fault clearing

7.1

Measurement check

The unit's measuring technology should be checked every 24 months. The following checks
should be performed in the process:
1. Check of the overall mechanical condition of the ergometer
2. Display check (contrast, lighting, etc.)
3. Check of correct ergometer rotational speed display
4. Check of the measurement sensor for brake performance
5. Check of mechanical power loss of the ergometer's drive system
6. Electrical safety check
7. Check of the safety symbols and markings on the casing
8. Writing of an inspection report
These checks and any recalibrating work necessary should only be performed by
authorised and trained personnel with the special tools required for this purpose.
7.2

Cleaning the device

The surface of the casing can be cleaned with a soft cloth that is dry or moist. Commercially
available cleaning agents for household appliances can be used. The saddle and the handlebar
should be cleaned with a leatherette cleaner.
It is imperative to make sure that no water penetrates the device.
The device should never be cleaned with petrol, nitro cleaner or acetone.
7.3

Checking and setting the supply voltage

On delivery, the unit is set for the local supply voltage (110/115 ~ or 230/240 V ~). The current
voltage setting is recorded on the mains module. The power unit covering lid on the bottom of
the unit must be opened to convert the voltage. The voltage can then be changed on the power
unit on-board with a special voltage selector switch.
Voltage conversion should only be performed by authorised and trained personnel.
7.4

Changing a mains fuse

The fuse switch is located in the centre of the mains module. The lid can be prised out of its
lock-in position with the help of a small screwdriver. It can then be pulled out of the fuse well.
Two fuses are located in the fuse switch. After a continuity check, change the fuses if necessary.
Return the fuse switch to the well and press it into the lock-in position. If the fuse won't stay in
position, immediately inform the after-sales service.
Replace fuses only with other fuses of the same type with the same specification:
2 x 1.25 AT for 230 V, or 2 x 2.5 AT for 110 V.
Article no.: 2.520091 Rev.: a