4.3.7. Output DC current generation
The "OUTPUT" section of the function selector switch defines the possibility of generating
an output DC current with the instrument, considering the selectable measuring ranges 0-
20mA or 4-20mA. The instrument may operate in the following modes:
DC current source → DC current generation (see § 4.4.7).
Simulation → simulation of a transducer in a current loop with auxiliary power supply
(see § 4.4.8)
The positions of the rotary switch are the following:
mA → Output DC current, which may be selected as indicated in Table 2
Percentage value (% STEP)
0%
25%
50%
75%
100%
120%
125%
Table 2: Selectable values of output DC current
Output current adjustment is possible with the options:
➢ %STEP → setting of values 0%, 25%, 50%, 75%, 100%, 120%, 125% of the selected
range.
➢ Quick adjustment → setting of values 0%, 50%, 100% of the selected range.
➢ FINE → setting of custom values in the range 0 ÷ 24mA with resolution of 1µA.
→ Output DC current with automatic ramp as indicated in Table 3
mA
Ramp type
Description
Slow linear ramp
Quick linear ramp
Slow step ramp
Quick step ramp
Table 3: List of available ramps for output current
4.3.8. Loop Power and HART 250 modes
In the LOOP POWER function, the instrument is capable of generating an output voltage >
24V DC to supply an external transducer and directly measure the corresponding loop
current.
The HART 250 function allows setting a 250 internal resistor for loop measurement on
transducers operating with HART (Highway Addressable Remote Transducer) protocol.
Range 0-20mA
Range 4-20mA
0mA
5mA
10mA
15mA
20mA
24mA
Not available
Action
Passage from 0% →100% →0% in 40s
Passage from 0% →100% →0% in 20s
0% →100% →0% with ramps of 15s
0% →100% →0% with ramps of 5s
EN - 10
HT8100
Commentato [a2]: 4.4.8
4mA
8mA
12mA
16mA
20mA
Not available
24mA