4.3.5. STORE and RECALL modes

For each measuring function, except for the current generation function, it is possible to
save the displayed value in the instrument's memory and to recall the data saved on the
display at any time. The instrument allows saving max. 100 pieces of data in the memory.
Proceed as follows:
Data saving
1. Select the "STORE" symbol flashing on the display, using the 4-arrow selector located
on the front panel of the instrument.
2. Press the ENTER key to save the data in the memory. The number of the memory
location is immediately shown on the secondary display of the instrument.
Recalling on the display and deleting the internal memory
1. Select the "RECALL" symbol flashing on the display, using the 4-arrow selector located
on the front panel of the instrument.
2. Press the ENTER key. The display shows the value of the measure corresponding to
the last memory location used and the number of the location itself.
3. Use the up or down arrow key of the 4-arrow selector located on the front panel of the
instrument to select the desired location. Press and hold the arrow keys for more than
1 second in order to carry out a quick search.
4. Press the CANCEL key to exit the function.
5. Switch off the instrument, then switch on the instrument again while pressing and
holding the CANCEL key to delete the internal memory.

4.3.6. Activation/deactivation of the internal functions

The following actions can be activated by pressing and holding the function keys indicated
in Table 1 when switching on the instrument:
Key
Setting of the modes 0-20mA or 4-20mA of current generation used by
RANGE
default by the instrument (see § 4.3.7)
Deactivation of the Auto power off function. The message "APO Off" is
provided by the instrument and the indication "APO" disappears from
MODE
the display. The function is automatically re-activated upon the
following instrument start-up.
HFR
Display of the internal Firmware version of the instrument.
Activation/deactivation by default of the sound associated with the
pressure of the function keys. The messages "Beep On" or "Beep Off"
ENTER
appear on the display.
Clearing of the instrument's internal memory. The message "Clr" is
CANCEL
immediately shown on the display.
Activation of continuous backlighting. The message "Blt On" is
A-HOLD
immediately shown on the display. The function is automatically
disabled upon the following instrument start-up.
Complete deactivation of backlighting. The message "Blt Off" is
REL 
immediately shown on the display. The function is automatically
disabled upon the following instrument start-up.
Table 1: List of internal functions of the instrument