Injection transformers
You can use injection transformers or isolators with Bode 100 for measurements in active circuits like
control loops of power electronic systems. Make sure the isolation of the injector is safe to use for your
application.
WARNING
Death or severe injury due to hazardous voltage levels possible.
If the isolation of the injection transformer fails, the DC potential of the equipment under
test will be present at touchable parts of the Bode 100.
► Only use injection transformers, isolated for the application's maximum working
voltage and overvoltage.
► Always connect the ground terminal of the Bode 100 to the ground terminal in the
laboaratory, using a solid connection of at least 3.6 mm² cross-section and not
longer than 10 m.
For more information on how to configure probes correctly, how to perform a calibration / correction
and how to set up the Bode Analyzer Suite, please refer to the Bode 100, Bode 500 User Manual.