Herunterladen Inhalt Inhalt Diese Seite drucken

Approaching The Pressure Value (Decreasing Pressure) - WIKA CPB5000HP Betriebsanleitung

Höchstdruck kolbenmanometer
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Pressure Balance High-Pressure
CPB5000HP

3.2.5 Approaching the pressure value (decreasing pressure)

 Turn the spindle pump anticlockwise to release pressure in the system.
 If the pressure is close to the next test level, make the fine adjustment with the spindle wheel.
Caution:
The piston is lowered very quickly just before equilibrium is achieved.
Caution:
Do not remove masses completely from the piston/cylinder system under
pressure.
Caution:
Remove the masses individually and regulate the pressure gradually to the
same extent.
 Once the spindle pump is turned back completely counter-clockwise, only the initial pressure
remains in the system.
 Now the high-pressure shut-off valve can be opened.
Open the high-pressure shut-off valve only if the pressure in the system has
been relieved to initial pressure. For this purpose, always first turn the spindle
pump back completely counter-clockwise.
 The initial pressure can be further relieved by carefully opening the low-pressure shut-off valve.
 By completely opening the low-pressure shut-off valve, the initial pressure will completely escape
back into the tank.
 The system is now completely relieved
Even if you wish to relieve the system from a high pressure completely down to
zero immediately, the spindle pump must first be completely turned back
counter-clockwise. Then open the high-pressure shut-off valve, and
subsequently open the low-pressure shut-off valve. The initial pressure then
escapes completely into the tank. The system is now completely relieved.
WIKA Operating Instructions Pressure Balance High-Pressure Version 2.5