Commissioning, Operation; Mechanical Mounting - Wika Rtd Zusatzinformation

Vorschau ausblenden Andere Handbücher für RTD:
Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

2. Safety / 3. Commissioning, operation
Transmitter model (only for design with transmitter)
Year of manufacture
Approval-related data
Sensor symbol
ungrounded
grounded

3. Commissioning, operation

DANGER!
Danger to life from explosion
By using a measuring insert without a suitable connection head (case), an
explosion risk occurs which can cause fatalities.
Only use the measuring insert in the connection head designed for it.
DANGER!
Danger to life from missing instrument grounding
With missing or incorrect instrument grounding, there exists a risk of
dangerous voltages (caused, for example, by mechanical damage,
electrostatic charge or induction).
Ground thermometer!
Electrical screening may only be grounded at one end, and outside of the
Ex area.
Observe the special conditions (see chapter 4 "Special conditions for use (X conditions)").

3.1 Mechanical mounting

Mounting within metallic enclosures
The screen must be fully coupled to the equipotential bonding of the connected enclosure.
Mounting within non-metallic enclosures
All electrically conductive thermometer components within the hazardous area must be
provided with equipotential bonding.
WIKA additional information RTD and TC, non-incendive equipment (Ex n)
= ungrounded welded
= welded to the sheath (grounded)
Before mounting and commissioning the instrument, ensure you
read the operating instructions!
EN
7