Medela Thopaz+ Gebrauchsanweisung Seite 8

Vorschau ausblenden Andere Handbücher für Thopaz+:
Inhaltsverzeichnis
Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 54
Safety related tips
– The product is delivered non-sterile� Before first use, as well as after each use,
the product must be cleaned and disinfected according to a validated procedure�
– Never place Thopaz + with an attached 2 l canister on the docking station without
attaching the optional adapter docking station 079�0038 for Thopaz + 2 l canister first�
– The acoustic warning "Leak in system – 301" is deactivated for approximately 6 minutes
after Thopaz + is switched on�
– Every time a new patient is connected, it is recommended to carry out the functional
check�
– Before the first application in accordance with the intended use, an inspection
according to IEC 62353 (see Appendix B) is recommended to acquire the reference
values�
– Use sample port of tubing to drain a sample� Therefore refer to chapter 11 – Taking a
drainage sample�
– Medela recommends to use the Thopaz tubing that best fits to the catheter�
– Make sure the sample port faces the patient�
– If the battery is low, use Thopaz + with mains supply�
– The simethicone capsule requires 60 min of submersion for full effect�
– For most accurate fluid measurement place or keep Thopaz + on the docking station
(079�0037)�
10 |
Warnings and safety instructions