Thopaz + , docking station, standard holder, and holder – standard reprocessing
Thopaz + in case of spill on seal
In case of spill on release button, bed holder, docking station's push button standard holder,
and holder with rail
x – Dry disassembled components in washer-disinfector at 110 °C for at least 45 minutes�
x x x – If drying in the washer-disinfector is not possible or in case of residual moisture,
wipe external surfaces dry using a dry, soft, lint-free wipe, or carefully dry with medical
grade compressed air�
– Pay special attention to the dryness of hard to reach areas�
x x x – Visually inspect the device or disassembled components for any remaining soil
or disinfectant solution� If necessary, repeat the cleaning and disinfection�
– Visually inspect the device or disassembled components for damage� In case of any
damage to one or more parts, replace them with new ones�
x – Put spring back in its initial position�
– Then push the release button carefully back into its initial position (see picture)�
– Verify that the release button springs back in position if pushed down to verify correct
placement�
– Re-insert silicone inlays in the handle�
– Insert the axles pieces into Thopaz + ' back cover�
– Slide handle into recess in axles and carefully tighten the screws using a Philips screw
driver (see picture)�
– Verify that the bed holder is mounted securely and can be opened and closed�
– To reassemble the docking station, re-install the spring on the shaft of the push button
and re-insert the push button in slotted hole in the docking station� Next, hold the push
button down and re-install the clamp and finally tighten the screw�
x
– Always store device in a dry, clean, and dust free environment�
x x x – Reprocess the device before sending it in for service� If this is not possible or can only
be done in parts, the package shall indicate the potential biohazard� Local procedures
and guidelines apply�
General reprocessing guidelines
| 49