Customer Responsibilities and Site Requirements
Customer Responsibilities and Site Requirements
Customer Responsibilities and Site Requirements
Make sure your site meets the following prerequisites.
• The necessary laboratory or bench space is available.
• The environmental conditions for the lab:
• The power requirements related to the product (100 – 240 V AC, 50 – 60 Hz)
• The required operating supplies necessary for the product and installation
Condensation within the instrument
C A U T I O N
Condensation may damage the system electronics.
➔
➔
6
•
Dimensions of the instrument: 400 x 310 x 310 mm (width x depth x
height). Allow approximately 75 mm of clear space around the
instrument for air circulation and electrical connections.
•
Dimension of Laptop: 380 x260 mm (width x depth)
•
10 – 37 °C (50 – 98.6 °F) for DNA applications
•
10 – 33 °C (50 – 91.4 °F) for RNA and Protein applications
as specified in the Site Preparation Checklist.
as specified in the Site Preparation Checklist.
Do not store, ship or use the TapeStation under conditions where temperature
fluctuations could cause condensation within the instrument.
If your system was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.
Agilent 2200 TapeStation System - Site and Safety Manual