Erläuterungen nach NE107
Bezeichnung
I
C
S
M
F
OI/TSP-DE
Meldung
DeviceStatus
DIAG NO.
F
50
F
51
F
52
F
53
F
54
I
57
I
58
F
65
M
66
M
67
C
71
F
72
I
74
C
75
S
76
S
77
S
78
S
79
S
80
SensyTemp TSP
Diagnose
Device related (Electronics)
Sensor related Channel 2
Measurement failure
Short Circuit
Open Line
Sensor Range overrun
Sensor Range underrun
One point trim active
Two point trim active
Application related
Configuration failure
Redundant sensor on channel 1 not available
Redundant sensor on channel 2 not available
Re-Configuration in Progress
Application failure
Analog Output trim active
Analog Output is fixed/simulated
Application warning
Limit HIGH HIGH
Limit LOW LOW
Limit HIGH
Limit LOW
Beschreibung
OK or Information
Check Function
Off Specification
Maintenance Required
Failure
LCD-Anzeiger