Aesculap
ELAN 4 electro micro reciprocating saw GA837
7.4
Preparation before cleaning
Applied part
►
Prior to first mechanical cleaning/disinfection process: Assemble
ECCOS holder GB087R in a suitable tray (e.g. JF222R).
►
Place the product in the correct position in the ECCOS holder GB087R,
see Fig. D.
Tools
►
Thoroughly rinse (through) the product with running, cold water.
►
Place the product on a wire basket suitable for cleaning.
7.5
Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
►
Use cleaning and disinfecting agents according
to the manufacturer's instructions which
– are approved for plastic material and high-
grade steel,
– do not attack softeners (e.g. in silicone).
►
Do not use cleaning agents that contain ace-
tone.
►
Observe specifications regarding concentration,
temperature and exposure time.
►
Do not exceed the maximum temperature of
60 °C during chemical cleaning and/or disinfec-
tion.
►
Do not exceed the maximum temperature of
90 °C during thermal disinfection with DI water.
►
Dry the product for at least 10 minutes at a
maximum of 120 °C.
Note
The indicated drying time is a guide time only. It must be checked taking
into account the specific conditions (e.g. load) and if applicable adjusted.
8
®
Validated cleaning and disinfection procedure
Product
Validated procedure
Applied part
Manual cleaning with immer-
sion disinfection
Manual pre-cleaning with
brush and subsequent mechan-
ical alkaline cleaning and ther-
mal disinfecting
Tools
Manual pre-cleaning with
ultrasound and brush, and sub-
sequent mechanical alkaline
cleaning and thermal disinfect-
ing
Reference
see Manual clean-
ing with immer-
sion disinfection -
applied part
see Mechanical
cleaning/disin-
fecting with man-
ual pre-cleaning -
applied part
see Mechanical
cleaning/disin-
fecting with man-
ual pre-cleaning -
tools