10. Decommissioning

Danger of injury from media escaping
under pressure.
Disassemble in an unpressurised state.
-
WARNING
Depressurise the system.
-
Danger of injury from aggressive media.
Depending on the measured medium,
-
this may constitute a danger to the
operator.
ATTENTION
Wear suitable protective clothing, e.g.
-
gloves, goggles.

11. Maintenance

The device is, in principle, maintenance free. If necessary, the housing of the device may
be cleaned with a damp cloth and a non-aggressive cleaning solution while it is switched
off.
With certain media may, however, deposits or contamination may accumulate on the
diaphragm. The specification of appropriate maintenance intervals for inspection is
recommended in this case. Once the device has been properly decommissioned, the
diaphragm can normally be cleaned with a non-aggressive cleaning solution and a soft
brush or sponge. Care should be taken while doing so. If the diaphragm is covered in
limescale, decalcification by BD SENSORS is recommended. See the Servicing / Repair
section with regard to this.
Incorrect cleaning can result in irreparable damage to the measuring cell. For this
reason, you should never use sharp objects or compressed air to clean the diaphragm.