Temperature limits
Environment / storage
Media-contacting parts
Electrical protective measures
Short circuit protection
Reverse polarity protection
Mechanical strength
One-sided overload
Vibration
Shock
Materials
Pressure connection / flange
Standard
Option
Separating diaphragm
Standard
Option
Venting / drain valves
Sealing plugs
Standard
Option
Nuts / bolts
Standard
Option
Type plate
Housing
Standard
Option
Seals (media contacting)
Standard
Option
Filling liquid
Standard
Option
Explosion protection
Certification
AX18-DPT200
Maximum safety limits for intrinsi-
cally safe version
Certification
AX-18B-DPT200
Max. ambient temperature
Miscellaneous
Optional display
Configuration
Optional mounting bracket
Protection class
Mounting position
Weight
Current consumption
CE conformity
Connections
Electrical connections
Process connections
Standard
Option
Without display
: -40 ... 85 °C
With display
: -20 ... 65 °C
Silicone oil
: -40 ... 100 °C
Fluorinated oil
: -40 ... 100 °C
Permanent
No damage if connections are reversed, but no function either
Equivalent to the maximum static pressure of the differential pressure measuring cell
5 g RMS (25 ... 2000 Hz)
100 g / 1 ms
Stainless steel 304 / 1.4301
Stainless steel 316 / 1.4401
Stainless steel 316L / 1.4404
®
Hastelloy
C-276
Stainless steel 304 / 1.4301
Stainless steel 316 / 1.4401
Stainless steel 304 / 1.4301
Stainless steel 316 / 1.4401
Stainless steel 316 / 1.4401
Cast aluminium with epoxy coating (blue)
Housing made of stainless steel 304 / 1.4301
FKM (-30...250°C)
NBR (-40...125°C), PTFE (-180...250°C)
Silicone oil (-40...125°C)
Fluorinated oil (-40...125°C)
IBExU 14 ATEX 1273 X / IECEx IBE 16.0005X
II 1/2G Ex ia IIC T4 Ga/Gb
II 2D Ex ia IIIC T 85 °C Db
Pi = 660 mW, Ui = 28 V, Ii = 93 mA, Ci 29.7 nf, Li negligible
IBExU 15 ATEX 1110 X / IECEx IBE 16.0006X
zone 1: II 2G Ex db IIC T6 Gb
Ex ia: -40 ... 60 °C
Ex d: -40 ... 65 °C
Type
: LCD
Number of lines
: 2
Number of digits
: 8
Bar graph
: 0...100%
Rotatable
: Yes, in 90° steps and / or by rotating the electronics housing
- Zero point / span locally using 2 buttons
- Local configuration with optional display
®
- Full configuration via HART
Material: C-steel or stainless steel 304 / 1.4401
Weight: 0.45 kg (including nuts and bolts)
IP 67
Any
Differential pressure transmitter: approx. 3 kg (depending on the version used)
Max. 23 mA
EMC directive: 2004/108/EC
- Connector terminals in the terminal compartment (wire cross-sections up to 2.5mm²)
internal thread 1/4" - 18 NPT / fixing 7/16 UNF
internal thread 1/4" - 18 NPT / fixing M10
oval flange ½" NPT internal thread
adapter M20x1,5
(85°C when not in use)
(Info: +125°C for a short time, max. 30 min.)
(Info: +125°C for a short time, max. 30 min.)
in accordance with DIN EN 60068-2-6
in accordance with DIN EN 60068-2-27
EN
Others: on request
Others: on request
Others: on request
Others: on request
Others: on request
Others: on request
Others: on request
35

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis