Endress+Hauser PROline 93 II2G Serie Bedienungsanleitung Seite 48

Werbung

Verfügbare Sprachen

Verfügbare Sprachen


10
Procedure:
1. Transmitter: remove cover (a, Fig. 4) from the connection compartment.
2. Remove blank cover for the cable entries for channel 1 and channel 2.
3. Remove special cable entry (supplied with sensors). Run both sensor connecting
cables through the cover (b) of the cable gland and into the connection compart-
ment.
4. Position the cable fixing sleeves (c) of the sensor cables exactly next each other
(Detail C). Position the ground contact terminals (d) and screw them tight.
f
e
b
Fig. 5:
Connecting the sensor connecting cable
5. Spread the rubber seal (e) along the side slit using a suitable tool (e.g. a large
screwdriver) so that both sensor cables are clamped within. Push the rubber seal up
into the cable gland bracket (f). Close the cover of the cable gland (b) so that it is
tight.
6. Plug in the sensor cable connectors analogue to the arrangement in Fig. 4.
7. Transmitter: secure cover (a) on the connection compartment.
PROline Prosonic Flow 93 FOUNDATION Fieldbus
g
d
.
c
C
Endress+Hauser