Herunterladen Diese Seite drucken

Medela Invia Motion Gebrauchsanweisung Für Patienten Seite 20

Vorschau ausblenden Andere Handbücher für Invia Motion:

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 32
3. Check the display for one of the following visual symbols.
Visual
What
symbol
does this
displayed
symbol
mean?
The
canister
is full
Blinking!
+
acoustic
signal!
20
What do I have to do?
Mute the acoustic signal by pressing the
1. Wash hands and prepare clean paper towel.
2. Remove the Invia Motion NPWT pump from the
carrying case. Switch the Invia Motion NPWT
pump off by pressing and holding the
for 3 seconds. Press only once.
3. Position the clamp next to the quick-connector
and close the clamp on the pump tubing and
close the clamp on the dressing tubing.
Clamp open
Clamp closed
4. Disconnect the Invia Motion NPWT pump tubing
from the dressing tubing by pressing on the
sides of the quick-connector.
Have a paper towel or tissue available to cover
the tip of the used tubing to protect against
wound fluid leakage.
5. Detach the full canister/tubing set by rotating the
canister to the left.
Discard the used canister/tubing set according to
instructions provided by your healthcare profes-
sional. ➞
button.
button

Werbung