1. Use a size 1 Philips head screwdriver to undo the central screw
(Fig. 5). Remove the D-Ring and both posts from the device (Fig. 6).
2. Select a same sized D-Ring. Place the female post inside the D-Ring
(Fig. 8).
3. The default D-Ring placement is on the inside of the device. However,
the D-Rings can be moved to the outside of the frame, if preferred
(Fig. 9).
4. Place the D-Ring with the female post into the hole on the frame.
5. Place the male post through the outside of the frame and slide the two
together. We suggest using the longer male post on all Pro Sport
models or when the frame is too thick for the short male post.
6. Re-insert the central screw and fasten. Making sure not to over
tighten.
Device Removal
1. Ask the patient to sit down with the leg bent 80° with the foot flat on
the floor.
2. Unfasten all straps from the outer side of the device and attach the
straps back to themselves. This will make it easier to put the device on
next time and will also prolong the life of the straps.
Accessories and Replacement Parts
Please refer to the Össur catalog for a list of available replacement parts
or accessories.
USAGE
Refurbishing
The device should be regularly sent to Össur for refurbishing. The
frequency of which depends on amount of activity by patient. Your
healthcare professional should help determine frequency of service; the
recommended minimum frequency is yearly. Refurbishment is provided
at the cost of the customer. There is a usual turnaround time of three to
five working days, plus shipping time.
Cleaning and care
• Hand-wash using mild detergent and rinse thoroughly.
• Air dry.
Note: Do not machine-wash, tumble dry, iron, bleach, or wash with fabric
softener.
Note: If the device is used in salt water or chlorinated water, rinse with
fresh water and air dry.
Hinge
• Remove foreign materials (e.g., dirt or grass) and clean using fresh
water.
• Lubricate with silicone spray or dry graphite if needed.
DISPOSAL
The device and packaging must be disposed of in accordance with
respective local or national environmental regulations.
LIABILITY
Össur does not assume liability for the following:
• Device not maintained as instructed by the instructions for use.
• Device assembled with components from other manufacturers.
• Device used outside of recommended use condition, application, or
environment.