IDENTIFYING TEST LEADS
1‐ Test lead with a big test Clip.
2 ‐ High‐voltage test lead.
3 ‐ Normal test lead (For insulation resistance test, you can connect an alligator clip to this
test leads probe for easy use if necessary. See below figure).
SPECIFICATIONS
Accuracy is specified for a period of one year alter calibration and at 18°C to 28°C, with relative humidity <75%.
Accuracy specifications take the form of:
±([% of Reading]+[number of Least Significant Digits])
AC Voltage
Range
Resolution
600V
1V
Input Impedance: 4.5M
Frequency Range: 45Hz to 400Hz
Resistance
Range
Resolution
200
0,1
Open Circuit Voltage: about 2. 8V
Overload Protection: 250V
Continuity
Range
Introduction
The built‐in buzzer will
sound if the resistance
is less than 50 ohm.
Insulation Resistance
Rated Test Voltage
Output Voltage
Nominal Current
Display Ranges
Test Range
Min. Resolution
Accuracy
All manuals and user guides at all-guidesbox.com
Accuracy
± (1,5%+5)
Accuracy
±(1.0%+3)
Open Circuit Voltage:
about 2.8V
250V
500V
1000V
rated test voltage X (1 ± 10%)
1 mA
2.000M, 20.00M,
200.00M , 2000M
0.25M ‐2000M)
0001M
0.25M ‐ 200M : ± (3%+3)
200M ‐ 2000M: ±(5%+5)