PANEL DESCRIPTION
1. LCD
2. "TEST" button:
Be used in insulation resistance fest. You can press the Button to start Insulation resistance test, and the test
stops about 30 seconds later automatically. At any time, you can press this Button to exit the test.
3. "LOCK" key:
Press this key to lock the insulation resistance test. Alter you press "TEST" Button to start insulation resistance
test, you can press "LOCK" key to lock the test and the display will show the symbol.
Locking the test means that continuous test voltage is provided.
To exit locking state, just press "TEST" Button again.
4. "HOLD" key;
Hold the present reading on display when in ACV and ranges.
Hold the present data an display except the position of the decimal when in M ranges.
5. "
"key
Used to enable backlight function. About 15 seconds later, the backlight goes off automatically.
6. Rotary Switch
Used to select desired function and range as well as to turn on/off the meter.
7. "E" jack:
Input jack to be connected to the ground of the object to be tested
8. "G" jack:
Input jack (shielding terminal / ground terminal of ACV. test]
9. "ACV" jack
ACV. test Input jack
10. "L" jack:
Input Jack which is to be in connection with the test point of the circuit to be tested.
11. High Voltage Indicator
After you press "TEST" button to enable insulation resistance test, this indicator lights as the indication of high
voltage being outputted.