Instructions For Use | 7 - Feb - 2023 | Version: B
Complete stage IV: Final rinsing
1. Completely immerse the product in deionized water (20-30 °C / 68-86 °F) and
rinse all accessible surfaces in 3 rinses of at least 1 minute each.
2. During the rinsing, maneuver mobile components in all directions as far as
they will go 3 times.
3. Thoroughly rinse all surfaces of the product with hidden crevices or complex
geometries at least 3 times with deionized water. Use a disposable 50 ml
syringe.
4. Allow residual water to drip off for a sufficient length of time.
Complete stage V: Drying
1. Dry the product using a soft, lint-free wipe. Use medical-quality filtered
compressed air (pmax = 0.5 bar) to dry the areas you cannot reach with the
wipe.
2. Make sure the product is clean and undamaged. Repeat the cleaning and
disinfection process if necessary.
Visual inspection
If any damage is detected, set aside the product immediately.
4 Reprocessing
4.5 Manual Cleaning and Disinfection