4.2
Staff Qualifications
Instructions For Use | 7 - Feb - 2023 | Version: B
NOTICE
Unsuitable cleaning agents and process chemicals
Corrosion damage, premature aging, and visible material changes
Only use cleaning and disinfecting agents that are approved for the product
Only use process chemicals that are compatible with the device's materials
according to the chemical manufacturer's recommendations
Follow all the chemical manufacturer's application specifications regarding
temperature, concentration, and contact time
Do not use process chemicals that can cause stress cracking or brittleness
of plastic materials
Avoid contact with chloride-based solutions
If the products should come into contact with chloride-based solutions,
rinse thoroughly with deionized water and dry completely
NOTICE
Use of ultrasonic bath
Damage to the device
The product must not be exposed to the ultrasonic bath
In many countries, the qualifications required by personnel responsible for the re-
processing of medical devices are regulated by law.
In any event, the reprocessing of medical devices must always fall under the re-
sponsibility of qualified personnel who have the necessary knowledge and exper-
tise.
This knowledge and expertise can be acquired by completing further training in
this field or as a result of completing a dedicated qualification with practical expe-
rience, supplemented by appropriate further training measures, where required.
4 Reprocessing
4.1 Safety Notices
Page 50 of 62