Instructions For Use | 7 - Feb - 2023 | Version: B
Complete stage I: Cleaning
1. Pre-clean the camera head including cable and plug at the conclusion of the
surgery by wiping it with a lint-free and wetted, but not dripping wipe until no
more residue can be seen.
2. Completely immerse the product for 2 to 5 minutes in the cleaning solution
prepared as per manufacturer's instructions.
3. Make sure that the cleaning solution touches all accessible surfaces and that
the product remains immersed throughout the entire cleaning time. When
placing the product, ensure that all air escapes from the hidden crevices.
4. While the product is immersed in the cleaning solution, clean it using a soft,
clean, lint-free wipe or with a soft bristled brush. Wipe or brush for at least
1 minute or until no more residue can can be seen.
5. During the cleaning, maneuver mobile components of the completely immersed
product in all directions as far as they will go 3 times.
6. Thoroughly rinse all surfaces of the product with hidden crevices or complex
geometries at least 5 times with the cleaning solution. Use a disposable 50 ml
syringe.
Complete stage II: Intermediate rinsing
1. Completely immerse the product in drinking water (20-30 °C / 68-86 °F). Fully
rinse all accessible surfaces in 2 rinses of at least 1 minute each.
2. During the rinsing, maneuver mobile components in all directions as far as
they will go 3 times.
3. Thoroughly rinse all surfaces of the product with hidden crevices or complex
geometries at least 3 times with drinking water. Use a disposable 50 ml
syringe.
4. Allow residual water to drip off for a sufficient length of time.
Complete stage III: Disinfection
1. Completely immerse the product in the disinfectant solution for at least
12 minutes.
2. Make sure that all accessible surfaces remain immersed in the disinfectant
solution throughout the entire disinfecting time. Remove all adherent air
bubbles from the component surfaces.
3. Wipe or brush the product, while it is completely immersed in the cleaning
solution, using a soft, clean, lint-free wipe or with a soft bristled brush for at
least 1 minute.
4. During the cleaning, maneuver mobile components of the completely immersed
in the disinfectant solution product in all directions as far as they will go
3 times.
5. Thoroughly rinse all surfaces of the product with hidden crevices or complex
geometries at least 5 times with disinfectant solution. Use a disposable 50 ml
syringe.
4 Reprocessing
4.5 Manual Cleaning and Disinfection
Page 53 of 62