Liquiphant FTL33
7.3
Function test with test magnet
WARNING
L
Risk of injury!
‣
Ensure that no dangerous processes are activated in the system.
To perform a function test, hold the test magnet against the marking on the nameplate (for at
least 2 seconds). This inverts the current switching status and the yellow LED changes state.
When the magnet is removed, the switching status valid at that time is adopted.
22
Test magnet and marking
The test magnet is not included in the delivery and can be ordered as an optional
accessory. → 65
Endress+Hauser
Commissioning
A0020960
63