Liquiphant FTL33
5.1.8
Installation in vessels
If installed horizontally, pay attention to the position of the tuning fork to ensure that the
liquid can drip off easily.
The electrical connection, e.g. M12 connector, should be pointing down with the cable. This
can prevent moisture from penetrating.
20
Position of the fork in the case of horizontal installation in a vessel
5.1.9
Distance from wall
Ensure that there is sufficient distance between the expected buildup on the tank wall and the
fork. Recommended distance from wall ≥10 mm (0.39 in).
Endress+Hauser
Installation
A0021034
A0022272
51