Commissioning
7.1
Function check
Before commissioning your measuring point, ensure that the post-installation and post-
connection checks have been performed.
• "Post-installation check" checklist →  54
• "Post-connection check" checklist →  60
The function of the tuning fork can be easily tested by immersing the tuning fork in a
vessel containing water.
7.2
LED display
A
2
3
A
M12 connector, (cable without graphic)
B
Valve plug
Item
Function
Green LED (gn)
1
Lit
Yellow LED (ye)
2
Lit
Red LED (rd)
3
flashing
Lit
For the metallic housing cover (IP69K), there is no external signaling via LEDs. A
connecting cable with an M12 connector and LED display can be ordered as an accessory.
Endress+Hauser
1
4
1
2
3
Description
Device is operational
M12 connector
Indicates the sensor state: tuning fork is covered by liquid
Valve plug / cable
Indicates the switching state:
• MAX operating mode (overfill prevention): sensor is not covered by liquid
• MIN operating mode (dry running protection): the sensor is covered by liquid
Warning/maintenance required: error can be rectified, e.g. incorrect wiring
Fault/device failure: error cannot be rectified, e.g. electronic error
B
2
3
3
2
Commissioning
1
1
A0016856