Herunterladen Inhalt Inhalt Diese Seite drucken

Shortening The Cable; Extension Cable; Parallel Connection - HBM PW27P Montageanleitung

Vorschau ausblenden Andere Handbücher für PW27P:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Electrical connection
excitation voltage leads. However, there will be a voltage loss on the supply
leads due to the cable resistance that is still present and not compensated for
by the six‐wire configuration. A large part of this loss can be eliminated by a
calibration, however, the temperature‐dependent part remains.
Important
The TK
value given in the specifications for the transducer therefore does not
c
apply for the cable and transducer combination when connection is in a
four‐wire configuration, where the cable percentage must be added.
5.3

Shortening the cable

If the transducer is connected to an amplifier in a six‐wire configuration, the
transducer cable can be shortened as required, without adversely affecting the
measurement accuracy.
5.4

Extension cable

Only use shielded, low‐capacitance measurement cables for extending. Ensure
that connection is perfect, with a low contact resistance.
The cable of a six‐wire transducer can be extended with a cable of the same
type.
5.5

Parallel connection

Only load cells with an aligned output (nominal (rated) sensitivity and output
resistance) are suitable for parallel connection. This option is not available with
type PW27P load cells.
12
A03631_02_Y00_00 HBM: public
PW27P

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis