Operating temperature and relative air humidity:
For operating temperatures from 0 °C to 30 °C: relative air humidity less than
80 %,
For operating temperatures from 31 °C to 40 °C: relative air humidity less than
75 %,
For operating temperatures from 41 °C to 50 °C: relative air humidity less than
45 %,
-
Storage temperature: The BENNING CM 1-1/ 1-2 can be stored at
temperatures between - 20 °C and + 60 °C (air humidity of 0 to 80 %).
During storage, the battery should be removed.

7. Electricalspecifications

Note: The measuring accuracy is specified as the sum of:
-
a relative part of the measured value and
-
a number of digits (i.e. counting Stepps of the last digit).
This measuring accuracy applies for temperatures from 18 °C to 28 °C and a
relative air humidity less than 80 %.
7.1 DC voltage ranges (BENNING CM 1-2)
The input resistance is 10 MΩ.
Measuring
Resolution
range
200 V
600 V
7.2 AC voltage ranges (BENNING CM 1-2)
The input resistance is 10 MΩ in parallel 100 pF.
Measuring
Resolution
range
200 V
600 V
7.3 AC current ranges
BENNING CM 1-1
Measuring
Resolution
range
20 A
0.01 A
200 A
0.1 A
400 A
BENNING CM 1-2
Measuring
Resolution
range
40 A
0.1 A
200 A
0,1 A
400 A
The measured value is obtained by mean value rectification and is displayed
*1
as r.m.s. value. Its calibration is adapted to a sinusoidal curve.
The stated accuracy is specified for conductors that are centrally clamped
by means of the current clamp L (see figure 4 AC current measurement).
For conductors that are not centrally clamped, an additional error of 1 % of
the displayed value has to be considered.
05/ 2008
0.1 V
± (1.0 % of the measured value + 2 digits)
1 V
± (1.0 % of the measured value + 2 digits)
Measuring accuracy
0.1 V
± (1.5 % of the measured value + 5 digits)
1 V
± (1.5 % of the measured value + 5 digits)
Measuring accuracy
the frequency range 50 Hz - 60 Hz
± (3.0 % of the measured value + 5 digits)
± (2.0 % of the measured value + 5 digits)
1 A
± (2.0 % of the measured value + 5 digits)
Measuring accuracy
the frequency range 50 Hz - 60 Hz
± (1.9 % of the measured value + 10 digits)
± (1.9 % of the measured value + 5 digits)
1 A
± (1.9 % of the measured value + 5 digits)
BENNING CM 1‑1/ 1‑2
Measuring accuracy
within the frequency
*1
range 50 Hz - 500 Hz
within
*1
within
*1
Overload
protection
600 V
eff
600 V
eff
Overload
protection
600 V
eff
600 V
eff
Overload
protection
600 A
eff
600 A
eff
600 A
eff
Overload
protection
600 A
eff
600 A
eff
600 A
eff
11