Connection of the High Vacuum Flange

Installation of ISO-K Flange

For ISO-K flange connections, fix the two flanges with the clamps
(high resistance 12.9 Agilent IC 63250 DCMZ) as shown in the
following figure (a protective screen can optionally be used).
Figure 39
Use the required number of claw clamps:
Tighten the claw clamps with a torque of 35 Nm
312/328
10 for ISO 100 K flange