Herunterladen Inhalt Inhalt Diese Seite drucken

Agilent Technologies TwisTorr 404 FS Bedienungshandbuch Seite 278

Inhaltsverzeichnis

Werbung

17
Technical Information
Description of the TwisTorr 404 FS
If the thermistor reading exceeds 60 °C the pump fails and stops with
Error Code = 2. If the thermistor reading is below 0 °C, the pump
fails and stops with Error Code = 1 (see Controller Manual).
The pump can operate in any position and can be supported on the
high vacuum flange. The connection of the forevacuum on the side of
the pump is a KF 25 flange.
278/328
TwisTorr 404 FS User Manual / 87-901-048-01

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis