ALIGNMENT INSTRUCTIONS

Bench Alignment (Fig. 3)
Sagittal Plane
Determine vertical socket position in alignment jig.
Adjust to appropriate heel height.
Introduce appropriate socket angles flexion.
The load line should fall at the junction of the posterior and the middle
third portions as shown in Fig. 3.
Coronal Plane
Introduce appropriate socket abduction/adduction using alignment jig.
Determine appropriate internal or external socket rotation.
USAGE
Cleaning and care
Clean with a damp cloth and a mild soap. Dry with a cloth after cleaning.
Environmental Conditions
The device is Waterproof.
A Waterproof device can be used in a wet or humid environment and
submerged in up to 1-meter-deep fresh water for a maximum of 30
minutes. No contact with salt water or chlorinated water is allowed.
Dry with a cloth after contact with fresh water or humidity.
Clean with fresh water in case of accidental exposure to other liquids,
chemicals, sand, dust, or dirt and dry with a cloth.
MAINTENANCE
The device and the overall prosthesis should be examined by a healthcare
professional. Interval should be determined based on patient activity.
Noise from Foot
Noise may occur if sand or debris is present in device. In that case, the
healthcare professional should doff the foot, clean it with the help of
compressed air and replace the Spectra sock if it is damaged.
REPORT OF SERIOUS INCIDENT
Any serious incident in relation to the device must be reported to the
manufacturer and relevant authorities.
DISPOSAL
The device and packaging must be disposed of in accordance with
respective local or national environmental regulations.
LIABILITY
Össur does not assume liability for the following:
• Device not maintained as instructed by the instructions for use.
• Device assembled with components from other manufacturers.
• Device used outside of recommended use condition, application, or
environment.
Compliance
This device has been tested according to ISO 10328 standard to two
million load cycles.
Depending on patient activity this may correspond to 2-3 years of use.
9