13.2. Connecting the cabinet pressure gauge to the compressor
 Remove the plug (1) from the threads (2) on the
pneumatic block of the compressor. Connect the
cabinet pressure gauge hose to the threaded
fitting.
13.3. Connecting the condensate vessel
 Route the hose (1) through the opening in
the rear wall of the cabinet for compressor
models with a dryer and connect it to the
vessel.
13.4. Vacuum connector
(Fig. 9)
The suction pump is equipped with suction and discharge hoses. Hoses connected to the
inlet/outlet to and from the suction aggregate are routed along the rear wall of the cabinet to its
lower part. Suction pump hoses may be connected to the piping in the floor or routed through the
rear opening in the cabinet (10). Connect the suction hose to the device and the discharge hose to
the piping routed outside the operator's work area. A noise silencer can be added at the suction
unit outlet to reduce the noise level caused by air passing through the hose (see Chapter 8.2.
Accessories). A noise silencer with a bacteriological filter must be added if the air outlet from the
suction unit is vented to an indoor area.
Risk of biological contamination. The discharge hose from the suction pump
needs to be routed outside the operator's working area otherwise a silencer with
a bacteriological filter needs to be added.