Herunterladen Inhalt Inhalt Diese Seite drucken

Mettler Toledo SevenDirect SD20 Benutzerhandbuch Seite 18

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

#
Description
2
Duration
3
Temperature
4
Endpoint type
5
Sensor name
6
Sample ID
7
Range indicator
8
Function
: displays the duration of the measurement.
: displays the number of data points which are already recorded.
Only if Methods > Measure > Interval reading is active, you can tap to
switch between these 2 modes.
Displays the temperature.
Tap to perform a temperature check while the meter is in idle state.
: gets the temperature value from the sample automatically.
: taps to change the temperature value used for the current
measurement only.
Displays the information of the endpoint type.
If a user belonging to the Administrator user group is logged in, you can tap
to enter the active method to adjust settings for the endpoint type and the
stability criterion.
1. Endpoint type, defined in Methods > Measure > Endpoint type.
Automatic: stops the measurement automatically when the signal
is stable.
Manual: manually stops the measurement by tapping Manual
endpoint.
Timed: stops the measurement after the predefined time.
2. Stability Criterion (pH only), defined in Methods > Measure > Stability
criteria. Details see Stability criteria.
Strict
Standard
Fast
Displays the information of the selected sensor.
Tap to change to another sensor if no sensor is defined in the method and
no ISM sensor is connected.
1. pH electrode condition
Slope: 95-105% / Offset: ±(0-20) mV (Electrode is in good
condition)
Slope: 94-90% / Offset: ±(20-35) mV (Electrode needs
cleaning)
Slope: 89-85% / Offset: ±(>35) mV (Electrode is defective or too
old)
2. Informs about whether a calibration is due based on:
Methods > Calibration > Calibration reminder
Menu > Settings > Analysis settings > Calibration expiry action
3. Inform about whether a verification is due based on:
Methods > Verification > Verification reminder
Menu > Settings > Analysis settings > Verification expiry action
Displays the information of the sample ID.
Tap to set the sample ID manually.
Only visible for pH measurement mode.
The green area shows the range covered by the current calibration.
The bold vertical line will move to indicate the current measured value.
If Methods > Measure > Limits of the measurement are defined, the limits
are also indicated on the bar, marked as vertical lines.
SevenDirect™ SD20

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis