7. Fault and system messages

Display
Meaning
Low battery power, device will only
work for a short while longer
Battery flat
Mains operation without battery:
wrong voltage
SEnS
No sensor connected
Erro
Connected sensor or device faulty
- - - -
Logger data is being read by the
- - - -
serial interface
No
Battery flat
display or
strange
Mains operation without battery:
symbols,
wrong voltage or polarity
device is
not res-
System error
ponding
to input
Instrument faulty
Err.1
Measured value above allowable
range
Sensor faulty
Err.2
Measured value below allowable
range
Sensor defective
Err.4
Value is too low to be displayed, tare
is set
Err.7
System error
Err.9
Connected sensor or device faulty
Measured value is far above or
below allowable range
WIKA Operating Instructions Hand-Held Pressure Indicator
Remedy
Replace battery
Replace battery
Check power supply, change it if
necessary
Switch off device and connect
sensor
If second sensor available, check if
device is ok. Return defective
device/sensor to manufacturer for
repair
When the data transfer is completed
the device will automatically return
to normal measuring display, no
remedy necessary
Replace battery
Check power supply, change it if
necessary
Disconnect battery or power supply,
wait a while, re-connect
Return to manufacturer for repair
Check: Is the pressure outside the
sensor's range? -> measured value
too high!
Return to manufacturer for repair
Check: Is the pressure outside the
sensor's range? -> measured value
too low!
Return to manufacturer for repair
Check: is value below '-2000'
(check tare)?
Return to manufacturer for repair
If second sensor available, check if
device is ok. Return defective
device/sensor to manufacturer for
repair
Check: Is the pressure outside the
sensor's range?