Herunterladen Inhalt Inhalt Diese Seite drucken

P.off) Power Off Function; Adr.) Communication Function; Al.) Alarm; Cloc) Real Time Clock - WIKA CPH 6200 Betriebsanleitung

Hand-held druckmessgerat
Vorschau ausblenden Andere Handbücher für CPH 6200:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Hand-Held Pressure Indicator
CPH 6200 / CPH 6200-D / CPH 6200-Ex
GB
14
Operation of min/max-value memory in conjunction with the
averaging function:
If averaging is activated and 'slow' measuring rate is selected
(rAtE-Slo), the min-/max-memory value relates to the average
display value.
If averaging is activated and 'fast' measuring rate is selected
(rAtE-FASt or P.dEt) , the min-/max-memory value relates to the
internal measured values (pressure spikes can be detected).

2.5 (P.oFF) Power off function

If no key is pressed, and no serial communication occurs during
the interval set in configuration (P.off), the device will automatically
switch itself off to save battery power.
If P.oFF = oFF then the automatic power off is deactivated.
2.6 (Adr.) Serial communications address
This parameter has been included in preparation for future
operation with multi-point monitoring systems.

2.7 (AL.) Alarm

There are three possible settings: Alarm off (AL. oFF), on with
audio alarm (AL. on), on without audio alarm (AL. no.So).
The following conditions will prompt an alarm to be displayed, if
the alarm function is activated (on or no.So):
Value is below lower alarm limit (AL. Lo) or above upper alarm
limit (AL.Hi).
Sensor error (Sens Erro)
Low battery (bAt)
Fe 7: System error (always with sound)
When an alarm occurs, the alarm arrow flashes, and, during serial
communication, the 'PRIO'-flag is set within the transmitted
message.

2.8 (CLOC) Real time clock

The real time clock is needed for time-stamping each data point
recorded within the logger function. As a result, the clock settings
should be checked regularly.
If the battery has been replaced the appropriate menu 'CLOC' will
be accessed automatically.
WIKA Operating Instructions Hand-Held Pressure Indicator

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cph 6200-dCph 6200-ex

Inhaltsverzeichnis