Hand-Held Pressure Indicator
CPH 6200 / CPH 6200-D / CPH 6200-Ex
3. Measuring of water level - display unit [m]
In preparation.
4. Operation of logger
The device supports two different logger functions:
"Func-Stor": each time the ´Store/Quit´ key is pressed a data
"Func-CYCL": measurements will automatically be recorded at a
The logger records 3 measurement values for each interval:
Instantaneous or mean value (depending on logger setting, see
below), Min-Peak and Max-Peak.
To evaluate the data, WIKA's Data logger evaluation software
GSOFT (V 1.7 or higher) must be used. The software also allows
easy configuration and operation of the logger.
If the logger is selected (Func-Stor or Func CYCL) the hold
function is deactivated; the ´Store/Quit´ key is then used solely for
the operation of the logger functions.
Min- and Max-Peak are, respectively, the minimum and the
maximum measured values during the last measurement interval.
Therefore both the instantaneous pressure values and the
measured fluctuations can be better analysed.
4.1 "Func-Stor": Storing Discrete Measurements
In this mode, each time the ´Store/Quit´ key is pressed a measure-
ment and its time stamp will be recorded. The data recorded can
be viewed either on the display (an additional menu item "REAd
LoGG" is displayed when scrolling through the Configuration
menu) or through the serial interface and a PC running the GSOFT-
software.
WIKA Operating Instructions Hand-Held Pressure Indicator
point will be recorded.
defined interval, set within the logger menu 'CYCL'
until the logger is stopped or the logger memory
becomes full. Logging is started by pressing the
´Store/Quit´ key for 2 seconds.
GB
15