Guidance and manufacturer's declaration — electromagnetic immunity
Upper Arm Blood Pressure Monitor including AC-adapter is intended for use in the electromagnetic
environment specified below. The customer or the user of this Upper Arm Blood Pressure Monitor
including AC-adapter should assure that it is used in such
IMMUNITY TEST
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
Electrical fast tran-
sient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short interruptions
and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
Note: U T is the a.c. mains voltage prior to application of the test level
ENGLISH
IEC 60601 TEST
COMPLIANCE
LEVEL
LEVEL
±8 Kv contact
±8 Kv contact
±2 Kv, ±4 Kv,
±2 Kv, ±4 Kv,
±8 Kv
±8 Kv
±15 Kv air
±15 Kv air
±2 Kv for power
supply lines
±2 Kv for power
±1 Kv for input/
supply lines
output lines
±0,5 Kv, ±1 Kv line
to line
± 0,5, ± 1 kV
±0,5 Kv, ±1 Kv,
line to earth
±2 Kv line to earth
<5 % UT
<5 % UT
(>95 % dip in UT.)
(>95 % dip in UT.)
for 0.5 cycle
for 0.5 cycle
<5 % UT
<5 % UT
(>95 % dip in UT.)
(>95 % dip in UT.)
for 1 cycle
for 1 cycle
70 % UT
70 % UT
(30 % dip in UT.)
(30 % dip in UT.)
for 25/30 cycles
for 25/30 cycles
<5 % UT
<5 % UT
(>95 % dip in UT.)
(>95 % dip in UT.)
for 5/6 sec
for 5/6 sec
30 A/m
30 A/m
ELECTROMAGNETIC ENVIRON-
MENT— GUIDANCE
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidi-
ty should be at least 30 %
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
Mains power quality should be that of
a typical commercial or hospital envi-
ronment.
Mains power quality should be that of
a typical commercial or hospital envi-
ronment. If the user of the Upper Arm
Blood Pressure Monitor requires con-
tinued operation during power mains
interruptions, it is recommended that
the Upper Arm Blood Pressure Moni-
tor be powered from an uninterruptible
pow er supply or a battery.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.